Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Tegam 135A: (1 year) CAL VER /5500 DATE: 11-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.4 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 20 NUMBER OF LINES: 101 CONFIGURATION: Fluke 5500A ============================================================================= # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F V 1.002 ASK+ K X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Warm-up time: None 1.004 DISP Temperature: 0 - 50degC 1.004 DISP Humidity: <80% 1.005 HEAD INITIAL SETUP 1.006 DISP [32] WARNING 1.006 DISP HIGH VOLTAGE is used or exposed during the performance 1.006 DISP of this calibration. DEATH ON CONTACT may result if 1.006 DISP personnel fail to observe safety precautions. 1.007 DISP Turn on the UUT. 1.007 DISP *NOTICE* Replace battery if "BAT" is illuminated 1.008 PIC 135v 1.009 HEAD {DC VOLTAGE} 1.010 HEAD DC VOLTAGE: {2V Range} 1.011 5500 2 1.9000V 0.05% 0.0001U 2W 2.001 HEAD DC VOLTAGE: {20V Range} 2.002 5500 20 19.000V 0.05% 0.001U 2W 3.001 HEAD DC VOLTAGE: {200V Range} 3.002 5500 200 190.00V 0.1% 0.01U 2W 4.001 HEAD DC VOLTAGE: {2000V Range} 4.002 5500 2000 1000.0V 0.1% 0.1U 2W 5.001 HEAD {AC VOLTAGE} 5.002 HEAD AC VOLTAGE: {2V Range} 5.003 5500 2 1.0000V 1% 0.0015U 60H SI 2W 6.001 5500 2 1.0000V 1% 0.0015U 10kH SI 2W 7.001 5500 2 1.0000V 5% 0.0015U 20kH SI 2W 8.001 HEAD AC VOLTAGE: {20V Range} 8.002 5500 20 10.000V 1% 0.015U 500H SI 2W 9.001 HEAD AC VOLTAGE: {200V Range} 9.002 5500 200 100.00V 1% 0.15U 60H SI 2W 10.001 HEAD AC VOLTAGE: {2000V Range} 10.002 5500 2000 700.0V 1% 1.5U 60H SI 2W 11.001 HEAD {RESISTANCE} 11.002 HEAD RESISTANCE: {20 MOhm Range} 11.003 5500 20 19.000MZ 1% 0.01/ 2W 12.001 HEAD RESISTANCE: {2000 kOhm Range} 12.002 5500 2000 1900.0kZ 0.2% 0.01/ 2W 13.001 HEAD RESISTANCE: {200 kOhm Range} 13.002 5500 200 190.00kZ 0.2% 0.01/ 2W 14.001 HEAD RESISTANCE: {20 kOhm Range} 14.002 DISP Connect the 5500A to the UUT as follows: 14.002 DISP [32] 5500A AUX HI to UUT V/OHM 14.002 DISP [32] 5500A AUX LO to UUT COM 14.002 DISP [32] 5500A NORMAL HI to UUT V/OHM 14.002 DISP [32] 5500A NORMAL LO to UUT COM 14.002 DISP Connect the SENSE leads DIRECTLY to the UUT. Connect 14.002 DISP the NORMAL leads to the SENSE leads AT the UUT. 14.003 5500 20 19.000kZ 0.2% 0.01/ CW 15.001 HEAD RESISTANCE: {2 kOhm Range} 15.002 5500 2 1.9000kZ 0.25% 0.025/ CW 16.001 ASK- R 16.002 HEAD {DC LOW CURRENT} 16.003 DISP Set the Function Switch to DCA. 16.004 DISP Set the Range Switch to mA-A. 16.005 PIC 135lc 16.006 HEAD DC LOW CURRENT: {50mA Range} 16.007 5500 50 50.00mA 0.25% 0.02U 2W 17.001 HEAD {AC LOW CURRENT} 17.002 DISP Set the Function Switch to ACA. 17.003 HEAD AC LOW CURRENT: {50mA Range} 17.004 5500 50 50.00mA 1.5% 0.05U 60H SI 2W 18.001 HEAD {AC HIGH CURRENT} 18.002 PIC 135hc 18.003 HEAD AC HIGH CURRENT: {10A Range} 18.004 5500 10 10.00A 1.5% 0.05U 60H SI 2W 19.001 HEAD {DC HIGH CURRENT} 19.002 DISP Set the Function Switch to DCA. 19.003 HEAD DC HIGH CURRENT: {10A Range} 19.004 5500 10 10.00A 1% 0.02U 2W 20.001 END